Validation of new invasive digital hemoglobinometer for hemoglobin estimation as point-of-care device among pregnant women in a facility setting India
DOI:
https://doi.org/10.47203/IJCH.2024.v36i02.018Keywords:
Point-of-care testing, Anemia, Maternal Health, Diagnostic Accuracy, ScreeningAbstract
Background: Anemia poses a significant health challenge for pregnant women (PW), and accurate and timely diagnosis is essential for effective management. Objective: The objective was to assess the diagnostic validity of the new digital hemoglobinometer for hemoglobin against laboratory-based hematology autoanalyzer among PW. Methodology: The study was conducted at a secondary-level healthcare facility among 204 pregnant women to be sampled conveniently, their sociodemographic and iron intake data collected, and hemoglobin levels assessed using a digital hemoglobinometer (Device A) and a hematology analyzer. Specificity, sensitivity, PPV, NPV, diagnostic accuracy, and method agreement were evaluated via Bland-Altman analysis and Lin's concordance correlation coefficient. Results: The proportion of anemia using the Device A was 64.7% while the hematology analyzer reported a proportion of 52.9%. Device A showed a sensitivity of 97.22%, specificity of 80.30%, and diagnostic accuracy of 86.3%, with substantial agreement indicated by Cohen's kappa coefficient (kappa = 0.72) and the weighted kappa coefficient for different grades of anemia was 0.67. Bland-Altman analysis revealed a mean difference (bias) of -0.28 (± 0.5) between the two methods, with limits of agreement at -1.24 and 0.68. Lin's concordance correlation coefficient of absolute agreement was 0.91. Conclusion: The DH showed high sensitivity and diagnostic accuracy for anemia detection in PW, with substantial agreement with the hematology analyzer. It offers a convenient and rapid alternative for POC hemoglobin estimation in resource-constrained settings.
Downloads
References
National Family Health Survey [Internet]. [cited 2024 Apr 24]. Available from: http://rchiips.org/nfhs/
Stevens GA, Finucane MM, De-Regil LM, Paciorek CJ, Flaxman SR, Branca F, et al. Global, regional, and national trends in haemoglobin concentration and prevalence of total and severe anaemia in children and pregnant and non-pregnant women for 1995–2011: a systematic analysis of population-representative data. Lancet Glob Health. 2013;1(1):e16–25.
Daru J, Zamora J, Fernández-Félix BM, Vogel J, Oladapo OT, Morisaki N, et al. Risk of maternal mortality in women with severe anaemia during pregnancy and post partum: a multilevel analysis. The Lancet Global Health. 2018;6(5):e548–54.
Anemia in Clinical Practice-Definition and Classification: Does Hemoglobin Change With Aging? - PubMed [Internet]. [cited 2023 Sep 29]. Available from: https://pubmed.ncbi.nlm.nih.gov/26404438/
Clinical and Laboratory Standards Institute. H15A3E: Quantitative Hemoglobin Determination in Blood [Internet]. 2000 [cited 2019 Aug 7]. Available from: https://clsi.org/standards/products/hematology/documents/h15/
Yadav K, Jacob OM, Ahamed F, Mandal M, Kant S. Use of Point of Care Testing (POCT) in measurement of hemoglobin. Indian Journal of Community Health. 2018;30(1 (Supp)):72–9.
Nichols JH. Point of care testing. Clinics in Laboratory Medicine. 2007;27(4):893–908.
Ramaswamy G, Vohra K, Yadav K, Kaur R, Rai T, Jaiswal A, et al. Point-of-Care Testing Using Invasive and Non-Invasive Hemoglobinometers: Reliable and Valid Method for Estimation of Hemoglobin among Children 6–59 Months. Journal of Tropical Pediatrics. 2021;67(1):fmaa111.
Anemia Mukt Bharat [Internet]. [cited 2023 Sep 29]. Available from: https://anemiamuktbharat.info/
Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive hemoglobin and invasive point?of?care hemoglobin testing compared with a laboratory analyzer. Int Jnl Lab Hem. 2014;36(1):56–61.
Digital-Hemoglobin.pdf [Internet]. [cited 2023 Oct 4]. Available from: http://mokshitcorporation.com/wp-content/uploads/2021/05/Digital-Hemoglobin.pdf
Toppo M, Pal DK, Gour D, Melwani V, Dubey M, Mishra A. Comparison of Performance of Digital Hemoglobinometer over Automated Hematology Analyzer for Hemoglobin Estimation and Its user-friendliness among the Pregnant Women in Selected District Hospitals of Madhya Pradesh. Indian J Community Med. 2019;44(1):31–4.
Dubey A, Murti S. Validation of a hemoglobinometer for use in outdoor blood donation camps. Asian J Transfus Sci. 2016;10(2):164–5.
Downloads
Published
How to Cite
License
Copyright (c) 2024 Areeba khanam, Ankit Chandra, Thejas Achary, Kashish Vohra, Aravind M, Archana Singh, Kapil Yadav
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.